The core technology of static volumetric method specific surface area and pore size analyzer

domestic,

Specific surface area and pore size analyzer

The main application of two major test principles. The first is dynamic chromatography, also known as dynamic method; the second is static capacity method, as the name suggests is static. The static capacity method is currently the international mainstream testing technology, ISO

, ASTM

Also included in the standards committee, but also

Chinese national standard

Included.

The specific surface area and pore size analyzer of the static capacity method works as follows: under a low temperature (liquid nitrogen bath) condition, a certain amount of adsorbate gas is introduced into the sample tube, and the adsorption partial pressure is directly measured by controlling the equilibrium pressure in the sample tube. Obtaining the adsorption amount of the partial pressure point by the gas state equation; increasing the adsorption equilibrium pressure by gradually introducing the adsorbate gas to obtain an adsorption isotherm; and gradually reducing the adsorption equilibrium pressure by gradually extracting the adsorbate gas to obtain a desorption isotherm, The sample is measured in a fixed volume of the sample tube, and the adsorbate does not flow relative to the dynamic chromatography, so it is called the static volume method.

For adoption

Static capacity method

For the specific surface area and pore size analyzer, the key technical points are nothing more than the pressure sensor, the performance of the vacuum pump, the adjustment system of the liquid nitrogen surface, the sealing performance of the test chamber, and the test used. The performance of the tile, the temperature control of the experimental system, the overall stability and automation of the instrument. The following Jine Spectrum Technology will introduce these technical points to you one by one, and explain what kind of technology is a useful and reliable technology, which is the technology that works on the stability, accuracy and repeatability of instrument testing. I hope to help the majority of users, and do not be fooled by false blowing information during the purchase of specific surface area and aperture analyzer.

1.

Pressure Sensor

During the test, the cavity gas of the entire system is constantly changing, and the pressure of the cavity changes accordingly. This requires us to have a special accessory to detect the pressure changes caused by these gas changes, then we introduce the pressure sensor as an important accessory. The sensitivity, frequency response characteristics, stability, and measurement accuracy of the pressure sensor directly determine the quality of the pressure sensor. The use of pressure sensors from internationally renowned brands guarantees good performance and ensures accuracy in reading pressure.

In addition, the accuracy of the reading pressure can be guaranteed without using a well-known brand, and a pressure sensor suitable for a static specific surface area and aperture analyzer is also selected. Pressure sensors are classified into different types, such as resistance strain gauge pressure sensors, semiconductor strain gauge pressure sensors, piezoresistive pressure sensors, inductive pressure sensors, capacitive pressure sensors, resonant pressure sensors, and the like. Different pressure sensors are available for different industries. Since the specific surface area and the aperture analyzer are evacuated during the test, the entire test chamber is in a low vacuum state. If the piezoresistive pressure sensor is selected, it is obviously inappropriate. The capacitive pressure sensor must be selected. . Capacitive pressure sensors are mainly used to measure low pressure and have high load capacity; while piezoresistive pressure sensors are suitable for measuring pressure changes under high pressure conditions.

2.

Vacuum pump

A vacuum pump is a device or device that uses a mechanical, physical, chemical, or physicochemical method to evacuate a pumped container to obtain a vacuum. For the specific surface area and pore size analyzer of the static volumetric method, the good vacuum of the overall test chamber can ensure the accuracy of the experimental data. Therefore, a reliable vacuum pump is very important for gas adsorbers such as specific surface area and pore size analyzers!

From the current market point of view, there is still a certain gap between the vacuum pump produced in China and the imported vacuum pump, regardless of the internal rotary vane quality and detail design of the vacuum pump, the noise level of the pump, the vibration amplitude or the oil leakage from the pump. The handling of instant noodles has to be improved. If a domestic vacuum pump is provided for the static volumetric specific surface area and pore size analyzer, high temperature and high pressure gas adsorption instrument, etc., the vacuum degree of the instrument test chamber will be affected, so that the reliability of the final data has to be questioned.

Some customers will make assumptions. If the domestic vacuum pump of the specific surface area and pore size analyzer, high temperature and high pressure gas adsorption instrument is also replaced by the imported vacuum pump, the accuracy of the test data should be credible? If you think so, it is a big mistake. First of all, if the vacuum pump is replaced, it needs to have a running-in period with the specific surface area and pore size analyzer, high temperature and high pressure gas adsorption instrument, and the direct replacement will cause the other data of the instrument to deviate from the data when the domestic vacuum pump is equipped. During this period, continuous debugging and technical improvement are required.

For the vacuum pump built-in or externally good, many customers have always held different attitudes, which is mainly related to the propaganda of the specific surface area and pore size analyzer, high temperature and high pressure gas adsorption instrument. When the vacuum pump is placed outside the instrument, the automation requirements will not be met. Manually turn on and off the vacuum pump manually during each experiment. The long-term placement of the vacuum pump in the air will also affect the life of the vacuum pump. Over time, the quality will also be affected. The exterior is not beautiful. If it is placed inside the instrument, it can solve the above problems, but the problems caused by the vibration and heat generated by the vacuum pump will affect the overall performance of the specific surface area and the analyzer of the aperture analyzer, high temperature and high pressure gas adsorption instrument. How to deal with these problems is a big technical problem for manufacturers who replace imported vacuum pumps and change vacuum pumps from external to built-in!

Jine Spectrum Technology has always used German imported vacuum pumps from beginning to end! Years of experience have shown that the performance difference between imported and domestic vacuum pumps is not small, and it also has a great impact on the performance of specific surface area and pore size analyzers, high temperature and high pressure gas adsorbers. In addition, Jine Spectrum Technology has always advocated that the instrument should not only be aesthetically pleasing but also highly automated. Therefore, Jine Spectrum Technology has built the vacuum pump into the gas adsorption instrument such as specific surface area and pore size analyzer, high temperature and high pressure gas adsorption instrument, and it is also very good. The problem caused by the vibration and heat dissipation of the vacuum pump is solved. This technology has always been a leader in domestic manufacturers!

3.

Liquid nitrogen surface conditioning system

In the process of testing the specific surface area and pore size using a static volumetric instrument, the sample tube was placed in a Dewar filled with liquid nitrogen. The portion of the sample tube below the liquid nitrogen surface is called the cold volume, and the portion above the liquid nitrogen surface is called the warm volume. As the experiment progresses, the liquid nitrogen will automatically evaporate continuously, so that the ratio of cold volume to warm volume will gradually change, requiring specific surface area and pore size analyzer to have special techniques to deal with the problem caused by liquid nitrogen volatilization. In order to keep the ratio of cold volume to warm volume unchanged, reliable and accurate data can be obtained. However, the amount of liquid nitrogen volatilization is undetermined by the tester. Therefore, the specific surface area and pore size analyzer are required to monitor the liquid nitrogen surface drop in real time to ensure that the cold volume ratio under the liquid nitrogen surface remains unchanged.

The specific surface area and pore size analyzer of Jine Spectrum Technology is equipped with a patented temperature probe system. When the liquid nitrogen surface drops, the temperature probe will monitor the temperature change, transmit the signal to the experimental software, and the software will issue commands to step. The height of the motor rises relative to the height to adjust the position of the cold volume of the sample tube, so that the cold volume remains unchanged to ensure the accuracy of the final specific surface area, pore size, pore size distribution, pore volume and other data.

Other domestic manufacturers claim to have a liquid nitrogen compensation mechanism, but according to customers, the so-called compensation mechanism is to regularly increase the liquid nitrogen surface, for example: control liquid nitrogen surface rises 2 mm every 2 minutes. Instead of real-time monitoring, the exact amount of liquid nitrogen surface rises. The data error of such tests is relatively large, especially for regions with different natural conditions. For example, in the south and north of China, professional laboratories and production plants, temperature and other environments have great differences, resulting in different levels of liquid nitrogen volatilization. If the liquid nitrogen surface is compensated according to the timing, the test data is difficult to convince. . When testing in a fixed area, the amount of compensation is adjusted based on existing experience. This is why the same sample is very good when tested by the instrument manufacturer; when the customer buys back the specific surface area and the pore size analysis instrument, the specific surface area, pore size, pore volume and other data are not ideal when testing in their own area. .

4.

Test chamber sealing performance

For the specific surface area and aperture analyzer using the static capacity method, if the test chamber is not sufficiently sealed, the air leakage is extremely likely to occur, and the test data is unreliable. Therefore, the specific surface area and pore size analyzer test chamber sealing performance is also a very important parameter!

Jine Spectrum's specific surface area and pore size analyzer series uses a cartridge design. The so-called cartridge design is to install all the solenoid valves, sensors and other components involved in the instrument on a large assembly block, which can ensure a high degree of integration and greatly increase the ultimate vacuum. Moreover, the cartridge design ensures good temperature uniformity of the pipeline. In addition, this design is very convenient for the upgrade, maintenance and maintenance of the instrument.

If you choose the gold-based spectrum technology configuration of the molecular pump V-Sorb2800 TP

Series specific surface area and aperture analyzer

The sealing of the pipeline will be further improved. Because this series of instruments uses all-metal VCR sealing technology, the sealing performance reaches 1 X 10 -10 Pa*m 3 /s, while the limit of rubber seal is 1 X 10 -6 Pa*m 3 /s. In addition, it is also equipped with a brand pneumatic valve, which can be used with VCR sealing technology to achieve the desired effect, and achieve the best realm of “good horse with good saddle”.

5.

Use of dewar

One of the essential parts of the specific surface area and pore size analyzer using the static volume method is the Dewar. The quality of Dewar directly affects the stability of the test process and the accuracy of the sample data. Let us first understand the history of the Dewar.

Dewar was originally invented by the Scottish physicist and chemist James Dewar. In 1892, Dewar ordered Berg to blow a glass into a special glass bottle. This is a double-glazed container with two layers of glass walls coated with silver and then the air between the two walls is removed to create a vacuum. The silver on the two layers of the gallbladder prevents radiation from dissipating heat, and the vacuum prevents convection and conduction heat dissipation, so the temperature of the liquid contained in the bottle is not easily changed. Later, Berg made the outer shell of nickel to protect the fragile glass bottle. On January 20, 1893, Dewar announced the invention of a special cryostat - later known as Dewar.

The specific surface area and pore size analyzer is always placed in a low temperature liquid nitrogen bath during the experiment. The container for the liquid nitrogen bath (Dova) must meet the following hard conditions, otherwise it will easily affect the specific surface area. Testing of data such as pore size and pore volume.

1.

Quality is safe and reliable. Since the specific surface area and pore size data were tested, the Dewar was bathed with liquid nitrogen (or liquid argon, etc.) at minus 196 °C. If the selection of Dewar is not correct, or the raw materials do not meet the requirements of the specification, and the conditions for the freezing of nitrogen are not met, it is highly probable that the Dewar explosion will occur. Lighter accidents may occur after skin contact. The safety of the operator is endangered by objects such as debris generated during the explosion. Reports such as such incidents frequently appear in the headlines of the media (click to view:

Report one

)! Therefore, the quality of Dewar must be reliable and guaranteed!

2.

The insulation performance is better. The dewar is composed of inner and outer double bile. The outer surface of the inner liner is wrapped with a multi-layer thermal insulator. Once the reflective material and the thermal insulation material fall off, the adjacent reflective material will directly contact the outer bladder, so that the liquid nitrogen of the inner liner is rapidly Expand and gasify until an explosion occurs. In addition, the liquid nitrogen bath placed in the Dewar with poor thermal insulation has a very fast evaporation rate, which accelerates the consumption of liquid nitrogen and additionally increases the experimental cost.

3.

Can support test time for a long time. When testing the specific surface area, it is completed in one hour or two. However, when testing microporous materials, especially for samples with very rich micropore diameters, the test time can be up to 3 days. This requires that the Dewar flask capacity of the specific surface area and pore size analyzer must be large, the insulation performance must be good, and the test time that can be supported must be long! Otherwise, the liquid nitrogen consumption will be exhausted in the middle of the test, at which point the tester will have to retest. Not only is it time-consuming and labor-intensive, but it has to be manually guarded during the next test to prevent liquid nitrogen from running out.

Perhaps you will ask: Will there be any remedy for filling the liquid nitrogen bath in time when the liquid nitrogen is almost exhausted? This method is certainly feasible, but through our many special experiments, the following conclusions are drawn: (1) When the liquid nitrogen bath is filled in time, the adsorption data of adjacent P/P0 points tested at that time will be greatly affected, of course, The accuracy of the final data. (2) You will find that the Dewar lifting motor is at the top of the test station when filling the liquid nitrogen bath. At this time, adding a liquid nitrogen bath is difficult to operate, and liquid nitrogen splashing is prone to occur.

In summary, when purchasing the specific surface area and the aperture analyzer, it is absolutely impossible to use the dewar with glass material, which is not only vulnerable to external human factors, but also more likely to cause explosion due to the quality problem of the Dewar itself. occur. Moreover, the insulation performance of the Dewar must be good and the capacity must be large. If you need to test microporous materials, the support test time must be long.

The Dewar bottle used by Jine Spectrum Technology is imported from Japan; the inner and outer gallbladders are made of stainless steel and have good thermal insulation performance; the 4-liter large-capacity design can support test time of more than 72 hours, easily satisfying the ratio of microporous materials. Test of surface area, pore size distribution, pore volume and other data.

6.

Sample processing temperature control system

All of Jine Spectrum's range of instruments adhere to the fully automatic design concept and are consistently embodied in our equipment. The specific surface area and aperture analyzer of the static capacity method is no exception. The most obvious and intuitive embodiment is that the instrument has no buttons other than the instrument switch button. All operations are performed in the computer software. For example, in the sample processing station, when performing sample pretreatment, it is only necessary to load the sample, and then input the processing temperature and the processing time in the software until the sample processing is completed. The instrument has an integrated temperature module built in to fully automate the scheduled experiment. Compared with the peers, the use of external temperature control tables is much larger, and manual adjustments are needed to adjust the processing temperature and time. The degree of automation is greatly limited. In addition, the specific surface area and aperture analyzer of the static capacity method of Jine Spectrum Technology is unique in that it has a time preset function regardless of the sample processing station or the sample test station. The tester can preset the experiment start time to midnight and calculate the end time of the experiment. When you go to work the next day, you will be pleasantly surprised to find that the test has just been completed, greatly liberating human resources and greatly improving the test efficiency! All thanks to the fully automated instrument design! Yes, this function cannot be implemented for external temperature meter instruments.

If the above technologies are mature, such specific surface area and pore size analysis will be stable, fully automated, and the accuracy and repeatability of the test data will be quite reliable!

For more technical guidance on specific surface area and pore size analyzers, please call Golden Spectrum Technology at 400-888-2667 or visit We also provide

True density tester

,

Rigid foam opening ratio closed cell rate tester

,

High pressure gas adsorber

,

Gas adsorption analyzer

For scientific research equipment, welcome to consult.

Vehicle Diagnostic Cables

We make OBD connector with terminal by ourselves, soldering type and crimping type are both available. Also 12V and 24V type. OBD1, OB2, J1939, J1708, J1962, etc. Also molded by different type, straight type or right-angle type. The OBD connector cables used for Audi, Honda, Toyota, BWM, etc. We have wide range of materials source , also we can support customers to make a customized one to replace the original ones.